SOP-GMP



SOP-GMP Introduction.


This image represent SOP-GMP used in pharmaceutical industries.

-The SOP-GMP or sometimes referred to as Standard Operating Practices (SOP's) are used to lay out operations consistently and repeatedly in the same repetitive manner. SOP’s must be available for every task that is used in the manufacture or testing of a regulated product. SOP’s ensure that all processes and standard operation procedure are consistently replicated, so even when there are changes in personnel, organizations avoid inconsistencies and do not run unnecessary safety risks. 

Standard Operating Procedures or SOP's are written step-by-step procedure that quality control (QC), quality assurance (QA), and production units use in order to assure that the accuracy and precision of the original product development is maintained in the transformation from small trial and batch work to full scale production of the product. SOP-GMP is an essential part of the consistent replication of the tasks that are used to produce a regulated product to a pre-approved quality specification.

Standard Operating Procedures are active documents that are routinely reviewed and amended as production processes are repaired, altered or replaced.  detail written instructions describing specific steps to follow in all activities under defined conditions, they are used to accomplish standardization when performing specific functions and used to ensure approved methodologies are used wherever possible.  Having defined all the appropriate standard operating procedures and practices they are then used as audit standards to ensure that these define methods are being meticulously followed and adhered to.


Regulatory Importance of SOP-GMP's.

This image portrays the use of SOP-GMP as used in industry.

It can be deduced from what is written above that standard operating procedures  forms the backbone of regulatory compliance activities in all companies regulated by Good Manufacturing Practices (GMP) rules and regulations.

Having described the use and importance ascribed to Standard Operating Procedures it is incongruous to now inform you that the number one failure in compliance is; “not following written instructions”.  This is closely followed by the number two failure: “no documented instructions available”.

It is hard for any validation professional to understand the mentality of persons that would patiently wait for a regulatory visitation – knowing it was routine for their operators and laboratory technicians to execute tasks without using a documented standard operating practice document.

It therefore becomes pretty obvious and very essential for personnel to be trained in the use of documented "standard operating practice" documents in the execution of their daily tasks.  Thus demonstrating that they are aware of why and how these SOP-GMP activities play such an important role in fulfilling the specific company’s regulatory requirements from WHO, FDA, EMEA or other national health authorities. Health authorities world-wide expect pharmaceutical, cosmetic and food producers to use approved manufacturing processes in written SOP-GMP format.


Common SOP-GMP Revelations.

1. Failure of management with executive responsibility to review the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency according to established standard operating procedures to ensure that the quality system satisfies the requirements of Part 820 and the manufacturer's established quality policy and objectives, as required by 21 CFR 820.20(c). Specifically, there is no evidence that SOP-GMP related management reviews have been conducted.

2. Failure to establish procedure for, and to conduct, quality audits to assure that your firm's quality system is in compliance with the established quality system requirements, and to determine the effectiveness of the quality system, as required by 21 CFR 820.22. Specifically, there is no evidence that you have established quality audit standard operating procedures or that quality audits have been performed.

3. Failure to ensure that personnel are adequately trained to perform their assigned functions and to document training, as required by 21 CFR 820.25(b). Specifically, employees involved in the manufacturing, packaging, testing, shipping, and complaint handling of in vitro reagents have not been trained in current Good Manufacturing Practice or the use of standard operating procedures; i.e. SOP-GMP. It was noted; you were able to provide documentation of any          type of training for only one of your current employees.

4. Failure to establish and follow document approval, distribution and change procedure, as required by 21 CFR 820.40. For example, standard operating procedures on how to control approval, distribution, and changes to device master records (DMRs), specifications, device history records (DHRs) and labeling for your strep A and hCG in vitro diagnostics were not available and documents not signed as approved for use were being used in product manufacturing and finished product testing locations during this inspection.

5. Failure to establish standard operating procedures to ensure that all purchased or otherwise received product and services conform to specified requirements, as required by 21 CFR 820.50. For example, quality requirements to be met by your contract manufacturer of your strep A and hCG in vitro diagnostics have not been defined or documented.

6. Failure to develop, conduct, control, and monitor production processes to ensure that your devices conform to their specifications, as required by 21 CFR 820.70. For example, you do not have any cleaning standard operating procedures for the equipment used in the production of strep A and hCG in vitro diagnostics.

7. Failure to ensure that all inspection, measuring, and test equipment, including mechanical, automated, or electronic inspection and test equipment, are suitable for their intended purposes and are capable of producing valid results, as required by 21 CFR 820.72 (a). Your firm has no approved standard operating procedures to ensure that equipment is routinely calibrated, inspected, checked and maintained. For example, procedures for calibration, inspection, checks and maintenance of your firm's digital scales, autoclave and spectrophotometer used in the manufacture and testing of your strep A and hCG in vitro diagnostics were not available.

8. Failure to validate with a high degree of assurance and approve according to established standard operating procedures a process whose results cannot be fully verified by subsequent inspection and test, and failure to document such validation activities and results, as required by 21 CFR 820.75(a). For example, there is no standard operating procedures in place for validation of your firm's manufacturing process, cleaning operations, software calculations or filling operations for your strep A and hCG in vitro diagnostics.

9. Failure to use a documented standard operating procedures for incoming product inspection and acceptance purposes, as required by 21 CFR 820.80(b). For example, there is no documentation of acceptance activities for incoming reagents, reagent bottles, bottle droppers, bottle caps, or labeling for your strep A and hCG in vitro diagnostics.

10. Failure to establish and maintain standard operating procedures for finished device acceptance to ensure that each production run, lot, or batch of finished devices meets acceptance criteria, and to hold finished devices in quarantine until all activities required by the DMR are completed, as required by 21 CFR 820.80(d). Specifically, entire lots of your strep A and hCG in vitro diagnostics were tested and released after the first day of assembly/production before all activities required in the DMR had been completed. For example, the following lots were released prior to completion of all activities required in the DMR: Mainline Confirms Strep A, lot 94610; Mainline confirms DOTS Strep A, lot 92720; Mainline confirms III hCG serum/urine, lot 96380 and Mainline Maxie pregnancy test urine hCG, lot 418F30.

11. Failure to establish and maintain standard operating procedures for implementing corrective and preventive action, as required by 21 CFR 820.100. Specifically, no corrective and preventive action procedures were available during this inspection.

12. Failure to establish and maintain SOP-GMP procedures to control labeling activities to assure that a designated individual has examined labeling for accuracy including, where applicable, the correct expiration date, control number, storage instructions, handling instructions, and any additional processing instructions prior to labeling release for storage or use, as required by 21 CFR 820.120(b). For example, the DHR does not include a date or signature of a person who examined and approved the expiration date and lot numbers on your labels for your Strep A and hCG in vitro diagnostics.

13. Failure to use documented standard operating procedures for verification of the label and labeling used for each production lot in the device history record, as required by 21 CFR 820.120(d). For example, the lot code and expiration labeling for the Strep A and hCG pregnancy screening test kits are not recorded in the DHRs.

14. Failure to maintain DMRs that include, or refer to the location of, device specifications, as required by 21 CFR 820.181(a). For example, the DMRs for your strep A and hCG in vitro diagnostics do not include or refer to the locations of all specifications for all reagents, bottles, dropper tip caps or dropper tips.

15. A serious SOP-GMP failure in not establishing or maintaining a design history file (DHF) for each type of device, as required by 21 CFR 820.30(j). For example, you have failed to provide a DHF for any of the devices that your firm manufactures.

16. Failure to maintain complaint files, and to establish and maintain standard operating  procedures for receiving, reviewing, and evaluating complaints by a formally designated unit, as required by 21 CFR 820.198. For example, you stated that you stopped documenting complaints over a year ago, and there is no evidence that complaints 02-05-037, 02-06-004 and 02- 6-023 were reviewed or evaluated.



Make certain that your validation is up to regulatory compliance requirements by ensuring that this complete chain is in place.


VPURSDQ - VRAIQOQPQ.


SOP-GMP.


News for Today


All documents are now available in any language. Simply, when you make a purchase; you will be issued with an order number. Email us this order number and the language required, and we will then translate the documents listed in your purchase order into the language you have chosen, and email them out to you. No extra charge.
To ensure your protocols are going to be cGMP compliant always ensure that the templates you use; are technically fully detailed, and use a comprehensive and interactive lead-through SOP.
Complex equipment will always require a full suite of protocols; however simple equipment can often be validated with a more basic combination iq/oq/pq template: Your Risk Assessment must define this requirement.
If assistance is required in selecting the appropriate documents then please make contact with us at email address: rc@validation-online.net (all lower case) and we will contact you. for urgent requirements ring direct +44 1842 820906.
Complete validation Equip/HVAC/Com/soft/Util/Steam Packages VMP/VRA/URS/DQ/IQ/OQ/PQ from $199.00 The ever popular Combination IQ/OQ/PQ FOR $199.00 The only one on the net - Steam Qualification package $199.00 These templates cut Authoring costs - 50/75% Reviewing time - 50% and execution costs 20/40%. The only interactive and fully detailed suit of protocols available on line for Autoclave and Steam cGMP validation.VP/URS/IQ/OQ/PQ Complete with integral LEAD-ME-THRU' SOP. The only cGMP compliant Risk Assessment Template available on line.


User Requirements Specification Template
$115.00

The User Requirements Specification template is the document that sets the standard, and specifies your requirements in a manner that ensures when a system or piece of equipment is selected for cGMP use that all essential support elements; i.e. maintenance, parts, operator & maintenance training, are planned and budgeted for.  It uses three levels of User Requirements Specification Template (URS), URS Level 1, 2 and 3, and is the only URS to guarantee traceability from the URS through to the final OQ and PQ functionality testing. A requirement mandated by cGMP regulations. It can be used on mechanical, electrical and software controlled, monitored or driven systems. 



Corporate Validation Manual $1,160.00

This definitive 1000 + page (including all attachments) Corporate Validation Manual arrives with you in USB memory stick format, this enables you at any time to download protocol or test-scrips documents and quickly edit them into company bespoke documents. In fact there are over $3,500.00 worth of superb documents, that form attachments to the DVM manual, which can be instantly copied. Once copied, the unique document interactive editing, allows you to produce high quality bespoke company documents; Such as the Validation Plan Template or the equally ubiquitous User requirements Specification template Design Qualification (DQ); in a few hours. The cost of the Definitive Validation Manual, will be recouped in the first few weeks of use. It will then go on to show a massive return on your original investment.